Chapter 6 System Commands
© 2010 Colby Instruments. PDL-100A Operating and Programming Manual Page 51
NET NETWORK COMMAND
The NET command sets the network addresses and other
network settings for the instrument. Addresses are stored in
non-volatile memory.
NET [IP | GW | NM | PORT | DHCP ] arg
IP = XXX.YYY.ZZZ.AAA
Set IP address
GW = XXX.YYY.ZZZ.AAA
Set Gateway IP address
NM = XXX.YYY.ZZZ.AAA
Set Network Mask address
PORT = XXXX
Set Network Port #
DHCP = ON | OFF
Set DHCP to ON or OFF
NOTE: Network values are stored in non-volatile memory and
do not take effect until the next power cycle or *RST command
is performed.
NET IP 192.168.100.10
NET NM 255.255.0.0
NET PORT 5678
NET DHCP OFF
HP : OUTPUT 705;"NET DHCP ON"<NL>
NI : CWGPIB.Write “NET IP 192.168.100.11”
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com