73
CLAMS Hardware Manual 1012-0001 v1.01
DO machine wash
DO employ a dilute (10%) bleach solution
DO expose to temperatures less than 80°C
DO employ mild acetic acid
DO NOT employ ammonia or alcohol
DO NOT expose to temperatures above 80°C
DO NOT employ petroleum based cleaners
The component shown below is manufactured from PVC Plastic and Stainless Steel:
DO clean (especially tip) prior to experiment
DO hand wipe with soapy water
DO employ a dilute (10%) bleach solution
DO expose to temperatures less than 50°C
DO employ mild acetic acid
DO rinse and dry thoroughly
DO NOT machine wash
DO NOT submerge
DO NOT employ ammonia or alcohol
DO NOT expose to temperatures above 50°C
The component shown below is manufactured from Mixed Materials:
DO clean prior to each experiment
DO hand wipe with soapy water
DO employ a dilute (10%) bleach solution
DO NOT machine wash
DO NOT submerge
DO NOT employ ammonia or alcohol
WARNING
This component contains an electrical device (fan) that will be damaged if submerged, exposed
to high levels of moisture or condensation. Keep fan dry and free of debris when cleaning.