CLAMS Hardware Manual 1012-0001 v1.01
3.1.2 Universal Flow Controller
The Universal Flow Controller (UFC) provides ventilation for eight cages and one reference cage. It also selects
one of the air supply lines (channels) for gas sampling. The UFC requires an external ventilation pump,
matched in size for the required ventilation rate of all channels.
Universal Flow Controller Specifications:
Mass Flow Range: 0-3L/min (each channel)
Mass Flow Resolution: 0.0003L/min
Physical Size (WxHxD): 30cm x 20cm x 35cm
Physical Weight: 6kg
Power Requirements: 12VDC 3A, 6W