72
CLAMS Hardware Manual 1012-0001 v1.01
Open Carefully
The compressed spring will forcefully push the hopper free from the bowl. To avoid damage,
do not drop the hopper while opening.
5 Cleaning
Columbus Instruments employs several materials in the fabrication of CLAMS cages:
Polycarbonate Plastic
Acrylic Plastic
Polyvinylchloride Plastic (PVC)
Some materials tolerate aggressive cleaning agents and methods while others require special handling.
Denotes an acceptable agent or method
Denotes an unacceptable agent or method
The components shown below are manufactured from Acrylic and PVC Plastic:
DO hand wash in soapy water
DO employ a dilute (10%) bleach solution
DO expose to temperatures less than 50°C
DO employ mild acetic acid
DO NOT machine wash
DO NOT employ ammonia or alcohol
DO NOT expose to temperatures above 50°C
The components shown below are manufactured from Polycarbonate Plastic: