Overview
3-7
No.
Name
Explanation
1
Handle
For the hand grip.
2
Paper Hole
Paper HoleOpen the recorder casingput your finger into the hole to
touch the paper
3
Label
Product Information Label
4
Support
To support the machine stand on the table
5
Fan
Internal heat dissipation channel Air outlet
6
Battery
Compartment
Built-in rechargeable lithium battery.
1 Battery Compartment
Warning
Improper operation may cause the battery to be hot, ignited or exploded or it may lead to the
declination of the battery’s capacity. It is necessary to read the user manual carefully and the
warnings and cautions before using of the rechargeable lithium battery (hereinafter called
“Battery”).
Only authorized installation or maintenance engineer can open the battery compartment and
replace the battery; and the rechargeable lithium battery of the same mode provided by our
company should be used.
3.6 Right Panel of the Main Unit
Warning
The accessories connected to the analog and digital interfaces should be validated according to
the respective IEC standards (e.g. IEC 60950 for data processing equipment and IEC 60601-1
for medical equipment). Furthermore all configurations should comply with the valid version of
the IEC60601-1. Therefore anybody, who connects additional equipment to the signal input
connector or signal output connector to configure a medical system, must make sure that it
complies with the requirements of the valid version of the system standard IEC60601-1. If there
is any problem, consult us or your local agent.
In order to ensure the patient safety, the summation of leakage current should never exceed
leakage current limits while several other units are connected to the patients at the same time.