Operation Instructions
5-12
5.3.2 How to Print, Delete and Copy Patient Cases in RECALL Interface
Press the up and down key of key to select the patient case;
Press the left and right key of key to select Print, Delete, Delete all and USB copy;
Press the START/STOP key, and the corresponding prompt information will pop out on the
screen;
According to the prompt information, press the MENU or START/STOP key to confirm
whether to conduct Print, Delete, Delete all or USB copy.
Press RECALL to exit RECALL interface.
USB Connection is an optional function.
You are recommended to use the USB Flash Drive supplied or designated by us, like Kingston,
PNY, ADATA or Apacer, or it may fail to be recognized by the ECG machine or cause damage
to reduce the performance and safety of the machine.
Our ECG machine can only recognize the USB Flash Drive of FAT or FAT32 format. Please
format it to FAT or FAT32 before using your USB Flash Drive. FAT and FAT32 are
respectively available for the USB Flash Drive in a capacity of 0~2G and 2G~8G.
5.3.3 Connect ECG to computer
1) Use the network cable to connect the ECG machine to your computer.
If the ECG machine is connected to your computer via a switch, please use the straight network
cable; if it is directly connected to your computer, please use the cross network cable.
2) Set the firewall of your computer: Start → Setup → Control Panel → Network Connection → Local
Connection (or directly double click the local connection icon at the bottom right corner of the
desktop) → General (on the popup window “Local Connection Status”) → Attributes → Advanced →
Setup → Disable (on the window “Windows Firewall”) → Save.
3) Set the TCP/IP address of your computer: Follow the above steps → Attributes → Internet Protocol
(TCP/IP) (double click) → Advanced → Advanced TCP/IP Setup → IP Setup → IP Address → Add →
input the IP address. If the remote address of the ECG machine is 192.168.2.40, please input the IP
address 192.168.2.40 and subnet mask 250.250.250.0 on your computer as below: