System Settings
4-19
keyboard of the second ID, input the corresponding second ID, and click the
extraction key on the soft keyboard to obtain the patient information. After
the extraction succeeds, the corresponding information will be automatically
inputted to the relevant options in the patient information page.
[Close]It fails to obtain the patient information by the extraction key.
[DICOM AE]It defaults to ACME_STORE. It can change according to the
server side folder name with the patient information.
[FTP User]When selecting the [FTP transfer], it is necessary to set this
option. The system sets this option as TEST by default. User can also set as
required. The set FTP user name needs to be consistent with the relevant
setting of the FTP receiving software.
[FTP Password]: When selecting the [FTP transfer], it is necessary to set this
option. The system sets this option as COMEN by default. User can also set as
required. The set FTP Password needs to be consistent with the relevant
setting of the FTP receiving software.
[FTP Port]: When selecting the [FTP transfer], it is necessary to set this option.
The system sets this option as 21 by default. User can also set as required. The
set FTP port value must be more than 20, and the set FTP Port needs to be
consistent with the relevant setting of the FTP receiving software.
Note
The FTP receiving software includes two types, i.e., 32-bit and 64-bit.
User can select a suitable server according to their computers.
Steps of FTP transferring files: Under the condition that the
instrument is connected to the network normally, select the [Printer]to
[Close], the [Save Option] is on, select the [Transfer Type] to [FTP
Transfer], and meanwhile keep the server port, user and password of
the FTP receiving software consistent with the setting of these three
options on the instrument. User can transfer files after setting the