23
Main board is defective Replace the main board
Inflation
repeatedly but no
results
The type of the patient is wrong
Check out the type of the patient in the patient
management
There is air leakage in the
accessories
Check out the cuff ,the NIBP tube or the
interface socket
The pump or air valve defective Replace the pump or air valve
DC power module defective Replace DC power module
The NIBP values
are incorrect
The type of the patient is wrong
Check out the type of the patient in the patient
management
A continue NIBP measurement for
a long time
The measurement should start every 5 minutes
at least
The cuff or tube air leakage
1.There is leakage in the cuff or is defective
2. There is leakage in the tube
The NIBP module is defective Replace the NIBP module
The main board is defective Replace the main board
4.6.8 RESP Failures
Symptoms Possible Cause Corrective Action
NO RESP
data
RESP function is disabled Enable the ECG module in the module set up
ECG cable or electrode is defective Replace the ECG cable
The patient connection not good Reconnect the patient
The ECG Lead connected in wrong
position
Reconnect the electrodes
The connection between ECG module
and DC power supply board poorly
connected
Reinstall the ECG module
ECG module defective Replace the ECG module
Main board defective Replace the main board
4.6.9 TEMP Module Defective
Symptoms Possible Cause Corrective Action