EasyManua.ls Logo

Comen NC5 - Technical Alarm Messages

Default Icon
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
Software is incorrect
Use the incorrect software for patient
monitor
Fail to upgrade ”no system
tips”
USB-stick defective or can’t
be detected by patient monitor
Chang the usb-stick for testing
Software is damaged or
incorrect name
Check the software name or change the
software
USB socket is damaged or
internal cable defected
Check the software on another patient
monitor to confirm if the usb-stick and
software no problem, then check the
patient monitor usb socket and internal
cable
Main board is defective Replace the main board
4.6.13 Technical Alarm Messages
Please refer to the User manual.

Table of Contents