EasyManua.ls Logo

ConMed ExcaliburPLUS PC - Page 75

ConMed ExcaliburPLUS PC
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Jacks.
With
10
ohms
connected,
BVARM
should
be
+0.85
to
0.98
Vdc
and
with
150
ohms,
it
should
be
+1.89
to
2.18
Vdc.
BVARM
should
also
match
VARM
at
A3TP17
to
within
10
mV
dc.
If
this
test
passes,
then
the
trouble
is
most
likely
in
the
DAC
A3U19
or
A.R.M.
comparator
circuitry
around
A3U23.
Note
that
a
faulty
DAC
is
also
likely
to
cause
Err
5.X
Alarms.
If
the
VARM
vs.
resistance
test
fails,
check
the
de
current
source
A3VR2
and
A3U21D
by
connect-
ing
a
dc
milliammeter
from
A3TP17
to
ground.
The
meter
should
read
from
0.49
to
0.51
mA.
If
VARM
test
reads
near zero
and
the
de
current
is
low,
repeat
the
current
test
with
A3J2
discon-
nected
to
eliminate
a
possible
VARM
short
to
ground
in
the
A9W5
harness
or
on
the
A5
PWB
and
shorted
transistors
A5Q2
or
Q3.
With
the
Return
Electrode
Jacks
open-circuited,
check
the
waveforms
on
the
A.R.M.
oscillator
collectors,
A5Q2
and
Q3.
See
Figure
3.7.
They
should appear
as
half-wave
rectified
sine
waves
with
a
frequency
of
34.5
-
38.1
KHz.
If
the
fre-
quency
is
too
high,
A5C22,
23,
44
or
45
may
be
open.
If
the
circuit
is
not
oscillating
and
VARM
is
0.6
-
0.8
Vdc,
check
the
feedback
resistors
A5R7
and
R8,
transformer
AST4
for
opens
or
shorts
and
transistors
A5Q2
and
Q3
for
opens.
If
the
circuit
is
oscillating
at
the
correct
frequency
but
VARM
does
not
respond
to
resistance
changes,
check
the
circuitry
from
the
secondary
to
A5T4
to
the
Return
Electrode
jacks
for
shorts
or
open
circuits.
4.7.7
RE
Amplifier
Troubleshooting
The
first
step
in
trouble
shooting
a
Power
Amplifier
(PA)
is
to
remove
all
PA
collector
and
base
fuses
(check
the
fuses
as
you
remove
them
and
note
if
any
of
them
are
blown).
This
will
prevent
possible
secondary
failures
in
the
PA
due
to
overload.
Use
a
VOM
to
check
for
shorted
bipolar
power
transistors
(A4Q1-Q3,
Q6-A8),
power
MOS-
FETS
(Q4-Q5),
and
snubbing
networks
(A5D9-
11,
D14).
Check
to
see
that
diodes
(D1-3,
D6-8)
are
functional.
Check
TP9
for
approximately
+110
Vdc
and
U2-6
for
+12
Vdc.
alibur
Check
the
Base
Voltage
Generator
and
Waveform
Generator
for
proper
function
before
proceeding.
If
everything
checks
this
far,
enter
dIA
2
(See
sec-
tion
4.7.2)
and
select
Monopolar
Pure
Cut
from
the
front
panel.
Confirm
that
A4TP10
and
TP4
are
approximately
0
to
10V
signal
swings,
and
the
same
shape
as
A3TP10.
Turn
off
the
power
to
the
unit
and
replace
the
base
fuses.
Reenter
dIA
2
and
select
Monopolar
Pure Cut
from
the
front
panel.
Use
an
oscillo-
scope
on
the
drains
of
A4Q4
and
Q5
to
confirm
that
the
power
MOSFETs
are
switching
properly.
Check
the
voltage
on
the
cathode
of
D10
for
approximately
39
Vdc.
If
the
unit
has
passed
all
tests
to
this
point,
it
will
be
necessary
to
replace
the
collector
fuses
to
trou-
bleshoot
further.
Proceed
with
caution!
4.7.8.
Power
Amplifier
Transistor
Replacement
All
bipolar
power
transistors
A4Q1-Q3
and
Q6-
Q8
may
be
replaced
singly
without
replacing
the
entire
set.
Calibration
should
be
checked
after
transistor
replacement.
When
replacing
the
power
transistors:
1.
Be
sure
to
install
the
insulator
pad.
No
ther-
mal
compound
is
necessary,
but
the
mating
sur-
faces
of
the
transistor,
insulator
and
heatsink
should
be
clean.
Replace
any
insulator
which
is
torn,
punctured
or
dirty.
2.
Ensure
that
both
pins
of
the
transistor
are
properly
centered
in
the
socket before
setting
the
transistor.
3.
Tighten
the
screws
to
4-6
inch-pounds.
Excessive
torque
may
cause
the
insulator
pad
to
be
cut,
destroying
its
dielectric
strength
and
shorting
+110
Vde
to
ground.
4.7.9
Troubleshooting
Resets
Apparently
random
restarts
of
the
unit
are
gener-
ally
due
to
temporary
loss
of
power.
If
the
unit
is
the
only
machine
in
the
room
to
restart,
check
for
loose
connectors
in
power
supply
circuits,
broken
or
weakened
power
cords,
and
a
disconnected
or
broken
+5
volt
regulator
on
the
A3
assembly.
ra
4-21

Related product manuals