4.7.10
Interference
with
Other
Equipment
Most,
but
not
all
equipment,
for
use
in
the
oper-
ating
room
environment
is
designed
to
tolerate
the
interference
created
by
electrosurgical
genera-
tors.
If
other
equipment
does
show
interference,
check
that
equipment
for
broken
cable
shields,
faulty
connectors,
or
open
grounds.
Video
cam-
eras
are
particularly
prone
to
faulty
cable
shields
since
they
are
subject
to
flexing
and
sterilization
procedures.
The
manufacturer
of
the
equipment
may
be
able
to
provide
guidance
in
fixing
the
problem.
Improperly
terminated
video
lines
can
also
increase
susceptibility
to
interference
by
loading
the
video
signal
level.
As
an
example,
when
coaxi-
al
cable
is
used
to
route
video
signals
to
several
locations,
a
75
ohm
terminator
is
sometimes
used
at
each
location,
creating
an
extra
load
on
the
sys-
tem.
The
other
extreme
is
when
no
terminator
is
used.
When
long
runs
of
coax
are
used
to
multi-
ple
loads,
it
may
be
best
to
use
a
video
distribu-
tion
amp
to
solve
the
load
matching
problem.
Interference
problems
can
also
be
minimized
by
repositioning
equipment,
rerouting
cables,
power
cords,
or
using
a
different
power
outlet.
4.8
PARTS
ORDERING
INFORMATION
To obtain
replacement
parts
or
additional
infor-
mation
regarding
your
unit,
write
or
telephone
according
to
the
contact
information
as
listed
on
the
inside
front
cover
of
this
manual,
or
contact
your
CONMED
distributor.
4-22
To
ensure
prompt
service,
please
provide
the
fol-
lowing
information:
Model
Number
Serial
Number
Number
Reference
Designator
and
Description
of
Part
Aspen
Part
Number
(if
known)
Quantity
Desired
Mailing
or
Shipping
Address
Preferred
Shipping
Means
(if
any)
Purchase
Order
Number
(if
applicable)
Your
Name
If
you
are
returning
a
unit,
obtain
a
Return
Authorization
(R.A)
Number
from
CONMED
Technical
Services.
Please
mark
the
R.A.
number
on
the
outside
of
the
carton
for
prompt
service.
4.9
Parts
Lists,
PWB
Layouts
&
Schematics
The
following
section
contains
printed
wiring
board
layouts,
parts
lists
and
schematic
diagrams.
The
parts
lists
are
shown
on
the
same
page
as
the
corresponding
schematic
fold-out.
Listed
are
the
replaceable
parts
available
from
CONMED.
Many
of
the
more
common
parts
may
be
available
from
local
electronic
suppliers.