OPERATION MICRO-TEMP
®
LT, Model 749
OPERATION AND TECHNICAL MANUAL
Page 26 of 75
F. Inspect the power plug for bent or missing prongs.
G. Connect the detachable power cord to the unit.
WARNING
Do not by-pass ground on power cord plug. Electrical Hazards may result.
H. Insert the plug into a properly grounded receptacle.
I. Lay the localized heat therapy pad flat towards the unit.
J. If the pad is already filled, check that there are no leaks.
WARNING
Leaky pads or hoses should never be used. Water leaks present a risk of infection
and should be handled accordingly.
K. Connect the pad to the MICRO-TEMP
®
LT unit as described in Section (2-3.1), Step
(G).
L. Check that the pad is flat and the connecting hose is not twisted or pinched.
M. Place the pad on the patient’s body part.
NOTE: The maximum contact surface temperature is 41°C.
NOTE: If at any time, power is removed from the unit without actuating the power switch
(ON/OFF switch) a power fail indicator will activate. To stop the indicator, restore power or
turn the power switch to the “O” (OFF) position.
NOTE: In the case that power is lost and then restored to the unit, the unit will resume the
previous operation mode.
NOTE: If pad is placed above unit, disconnect pad prior to turning off unit. This could cause
backflow if pad is not disconnected.