4.9 Instrument handling
Follow the steps below for instrument handling:
(1) Release the brake of the instrument's casters;
(2) Hold the handle on the rear panel with both hands and push the instrument to a new
position. The pushing process cannot be too fast;
(3) When the instrument reaches the new position, lock the caster brakes.
Be sure to turn off the instrument before moving the device.
Before moving the instrument, the power cord must be disconnected from the main
power supply and wait for the instrument to cool down.
Before moving the instrument, make sure that there is no paraffin in the wax bath or
waste wax collection tray, and the wax boxes and activated carbon doors are closed to
prevent the wax bath from falling out during the movement. Or remove the wax bath
from the device.
During the movement process, the cylinder head is locked.
To avoid spilling or falling of reagent bottles due to shaking of reagents in the reagent
bottles during the movement of the equipment, all reagent bottles must be emptied or
taken out before moving.
During the moving process, ensure that the reagent compartment door is closed, and
at the same time use tape to fix the door to avoid opening the door during the moving
process.