9.2.2.4 Display Screen
Use a piece of lint-free cloth to clean the display screen. Click the clean
display screen button before cleaning to enter the screen clean state. During a
clean state, a short touch on the screen does not trigger any operation buttons.
After cleaning, press and hold the restore button in the lower right corner of
the screen. The interface exits the clean state.
9.2.3 Monthly Cleaning and Maintenance
9.2.3.1 Reagent bottle
Empty the reagent bottle. Use a brush and warm water for cleaning. After
drying, use corresponding reagent to flush the reagent bottle.
In the cleaning process, be careful not to damage the O-ring of the reagent
bottle and to avoid scratches on the surface of the seal ring.
The seal ring on the joint can be lubricated with the seal ring lubricant
provided with the machine, so as to ensure the sealing effect of the seal ring
and the smoothness of plugging and unplugging.
After cleaning, refill the reagent bottle. Screw the bottle cap. Insert the bottles
in the correct order into the reagent compartment. Check whether the blue
indicator light of the reagent bottle is on and check whether the liquid level of
each reagent bottle meets the requirements for the minimum liquid level.
9.2.3.2 Paraffin case
Before cleaning the paraffin case, TURN OFF the instrument and wear
personal protective equipment such as protective clothing, goggles, gas masks,
rubber gloves, etc.
The cleaning process requires careful operation to avoid burns caused by skin
contact with the interior or the periphery of the paraffin case.
Use a piece of lint-free cloth to wipe the interior of the paraffin case and the
metal plate surface of the paraffin case door.
Use a piece of lint-free cloth to wipe the seal ring of the paraffin case door.
9.2.4 Reagent Rinse
The following cleaning reagents are recommended for cleaning the instrument:
70% alcohol;
Paraffin removing agent (xylene or isopropanol);
Distilled water.