Dakewe (Shenzhen) Medical Equipment Co., Ltd
9.Cleaning and Maintenance
9.1 Rinse Program
The rinse program is a pre-installed program.
The rinse program cannot be edited, copied or deleted.
After the dehydration program is completed, the rinse program will be given
automatically. At this point, the dehydration cylinder must be cleaned. This function
cannot be cancelled.
Use only recommended rinse reagents.
Do not open the dehydration cylinder lid during the drying procedures of the rinse
program.
Do not pull out the reagent bottle during the operation of the rinse program. Such
operation may damage the instrument or trigger an error report.
The sample baskets can be cleaned together in the dehydration cylinder. Only the
sample baskets that have gone through standard rinse can be used for dehydration.
Xylene for rinse and alcohol for rinse are recommended to be replaced after 5 times of
rinse, otherwise it will affect the dehydration effect.
HP300 Plus includes three rinse processes: fast rinse, standard rinse and warm rinse
processes. Which are been used in the following situations:
Fast rinse: After running the dehydration process that does not include the paraffin
soaking step, the device can select fast rinse process for system cleaning to save
cleaning time.
Table 1: Fast Rinse Process (about 0.5h)
Positive and
negative pressure
drying
Vacuum +
positive pressure
Standard rinse: The default rinse process of the device. After performing a dehydration
process that includes the paraffin soaking step, the device is forced to run a standard
dehydration process.