7.Reagent Processing
7.1 Reagent Filling / Evacuation / Replacement
The following steps can only be carried out by trained laboratory personnel with reagent
handling experience.
When handling the reagents used in this instrument, be sure to wear personal protective
equipment and operate with caution.
Dispose of waste reagents carefully in accordance with local laws and regulations and
waste management policies of the Company or organization.
Do not change reagents or fill empty reagent bottles during program operation. This can
cause serious damage to the instrument.
Make sure to place the reagent bottle in the correct position according to the reagent
bottle layout on the display screen.
1. Remove the reagent bottle from the reagent case. Unscrew the bottle cap (Figure
66-1).
2. Thoroughly clean the reagent bottle before refilling.
3. The top and bottom filling liquid level are imprinted on the front of each reagent
bottle as a guide. During filling, use a funnel to keep it clean.
4. After filling, the cap needs to be screwed back to the bottle.
5. Put the filled reagent bottle back into the reagent case. When inserting a reagent
bottle, make sure that the reagent bottle joint’s seal ring is intact and accurately
inserted in the socket (Figure 66-2). A clicking sound indicates that you have
properly installed the reagent bottle in place.