62
Dakewe (Shenzhen) Medical Equipment Co., Ltd
6.2.1.5 Other settings
Temperature format (Figure 56-1): The system can be switched between Celsius and Fahrenheit.
Touch to activate the corresponding option.
Default setting the program end time (Figure 56-2): Touch to input, then click Save. When the
program is running with a delay start selected, the system will calculate the start time based on the
program’s default end time. For instance: The default program end time is set to 8 o’clock in the
morning, and the total running time of the program’s dehydration program is 10 hours, the device
will automatically start the dehydration program at 22 o’clock and end the program at 8 o’clock
the next morning.
Stirring time (Figure 56-3): Touch to input. The user’s dehydration program has enabled the
stirring function in some steps (see 5.3.1.1 Dehydration Program Creation). The system will stir
according to the set time of this option.
Complete reminder (Figure 56-4): This option allows the user to set the alarm time for the
completion of the program and the rinse program that is not executed. Click the triangle on the
right to pop up the drop-down box to select the desired duration to complete the setting.
Turn off the computer: See 4.8.5.