Dakewe (Shenzhen) Medical Equipment Co., Ltd
During the strainer removal, make sure that no foreign matters fall into the liquid inlet
and outlet at the bottom of the dehydration cylinder.
paraffin case door. Wipe it clean with a piece of lint-free cloth moistened with
xylene.
A paraffin filter is installed in the paraffin cylinder. If the filter surface has
more residual dirt, it can be disassembled for cleaning before being installed
in the paraffin cylinder.
Check the storage capacity of waste wax in the waste wax collection tray. Use
your fingers to hook the handle of the waste wax collection tray and pull out
until you see the end of the waste wax collection tray. Check whether there is
a large amount of waste wax inside, such as waste wax. Too much, you need
to remove the waste wax and insert the collection tray back into the storage
tank.
Due to high temperature inside and around the paraffin case, gloves shall be worn
during cleaning operation. Meanwhile, be careful to prevent burns.
9.2.2.2 Waste liquid collection bottle
The waste liquid collection bottle must be taken out to empty the waste liquid
inside and then be inserted back into the original place every week. Waste
liquid evacuation is prompted in the software. It is necessary to promptly
empty the waste liquid collection bottle after the prompt.
Due to the corrosiveness and toxicity of the liquid in the waste liquid collection bottle,
wear gloves while performing cleaning operation. Meanwhile, be careful to avoid being
contaminated by the waste liquid. Furthermore, waste liquid shall be disposed of
according to the local laws and regulations as well as the waste management policies of
the Company or organization.
9.2.2.3 Dehydration cylinder
Rotate the fixed bolt of the filter at the bottom of the dehydration cylinder.
Remove the strainer of the dehydration cylinder. Remove the solid dirt on the
strainer surface and in the filter holes. The strainer can be soaked in xylene or
alcohol for cleaning.
Check the seal ring of the dehydration cylinder lid regularly for damage. If the
seal ring is damaged, it must be replaced immediately. The airtightness of the
dehydration cylinder can be checked through the software component’s
quality control data. Meanwhile, the machine itself has the self-checking
function of the components. The leakage problem can be analyzed and
alarmed. If leakage is found, contact Dakewe Technical Service Department
for assistance.