26  Manual  RS.8A.N5.02  ©  Danfoss  10-2007  m2  Version 3.11
  Curve    This eld is used to select 1 of the 4 curve tables    
      (Cf1 - Cf4).
  Break    This eld is used to select the break point being    
      set up on the curve (1 - 12).
  Input    This eld is for the Internally Scaled input value to    
      be used at the break point selected. The 'To use ='    
      value can be copied into this eld by pressing ' '    
      and ' ' together. This eld has a range of between    
      0 and 9999.
  Output  This eld is for the Output value at the Break Point/   
      Input value selected. This has a range of between    
      -999.9 and 999.9.
The table should be completed so that the Input readings increase in value. Any 
reading below the rst value or above the last value will return out of range 
Output Values (OC/SC).
In order to set up a Table you will require either a Resistance box (for resistance 
probes) or a Current Source (for Current Probes) plus the manufacturer's data 
dening the Input/Output relationship of the probe. The method is described 
for a Resistance probe, but it is the same for the Current case except that it is 
necessary to put the Input 1 link in the 4-20 mA position before starting. 
1.  Select (up to) 12 Output (Resistance) values from the        
  manufacturer's data sheet which correspond to Input (e.g.      
  Temperature)  values in the range you wish to measure. For      
  maximum accuracy these points should be distributed evenly      
  across the required range.
2.  Connect the Resistance Box across the Input No. 1 and C      
  terminals of the unit and set it to the lowest of your selected      
  Output values. When you do this you should see a value appear    
  in the 'To use =' eld. This is the Internally Scaled version of the     
  resistance you have connected. Set the Break eld to 1 and then    
  go to the Input eld.
3.  Next either manually enter the value displayed in the 'To use ='     
  eld, or press the ‘ ‘ and ‘ ‘ keys together to copy it to the Input eld.
4.  Move to the Output eld and enter the Output (e.g. Temperature) value 
  from your probe data corresponding to the Resistance value you set.
5.  Move the Edit Box to  the Break eld again and increase its value by 1, 
  which will clear the Input and Output elds.
6.  Set your next chosen Resistance value from the probe data and,    
  when the 'To use =' value has stabilised, repeat  from step 3      
  until all your points have been entered.
Once all points have been entered, you can vary the resistance connected to 
Input 1 and observe the eect in the Converted eld. For a given resistance, this 
value should now be the same as that shown in the manufacturer's probe data.
For 4-20 mA inputs the 'Input type' must be set to 'CURRENT', then the Input/
Output elds will change to 4 mA/20 mA. The curve table can now be set up. 
For example, if a pressure sensor was being used which produced 4 mA at 0.1 
bar and 20 mA at 30 bar, then simply add 0.1 to the 4 mA eld and 30 to the 20 
mA led. Ensure the point set-up and jumper link on the input terminal is set to 
4-20 mA.
When you have nished or to abandon the operation, press ‘ ’ to return to 
the Setup Menu.
Setting up a Curve Table
4-20 mA Set-up