VER VERSION
VER [CMT]
Causes "version" information (Model No. and Software Version No.) to be transmitted
from the RS-232/485 Interface Port, followed by [EOT].
ZRO ZERO
ZRO=v [CMT]
Sets the ZERO OFFSET ("b" term) so that the existing input yields a reading equal to v
(-32700 v 32700). In "TWO-POINT" calibration, must precede the FORCE (FRC)
command. For RS-485 (only), returns ACK [EOT].
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION— THAT IS,
WHEN THE CAL=MXB [CMT] COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT], UNPREDICTABLE READINGS COULD OCCUR.
[NOTE: There is no valid interrogation form for the ZRO command.]
B.11
COMMAND AND RESPONSE SYNTAX
B
B.8 3500 MNEMONIC COMMANDS