Introduction
ELITEpro XC SAFETY SUMMARY and SPECIFICATIONS
4RH8
E186827
Conforms to UL Std 61010-1
Certified to CSA Std C22.2 No.
61010-1
The ELITEpro XC is an Over-Voltage Category III device. Use approved rubber gloves with mechanical
protection and goggles when operating the device.
CAUTION: This LOGGER may contain life-threatening voltages. QUALIFIED PERSONNEL MUST disconnect all high
voltage wiring before using or servicing the LOGGER.
Warning: Use of this device in a manner for which it is not intended may impair
its means of protection.
SYMBOLS ON EQUIPMENT
Denotes caution. See manual for a description of the meanings.
When connecting the ELITEpro XC to an AC load, follow these steps in sequence
to prevent a shock hazard.
1. If possible, de-energize the circuit to be monitored.
2. Connect the CTs to the phases being monitored.
3. Connect the voltage leads to the different phases. Use proper safety
equipment (gloves, mask, and protective clothing) as required for the
voltages monitored.
Risk of electric shock. Life-threatening voltages may be present. Qualified
personnel only.
DO NOT EXCEED 600V Phase to Phase. This logger is equipped to monitor loads
up to 600V. Exceeding this voltage will cause damage to the logger and danger to
the user. Always use a Potential Transformer (PT) for loads in excess of 600V. The
ELITEpro XC is a 600 volt Over-Voltage Category III device.
Dispose of properly.