0065-0647-OM Rev E 7550 OWNER’S MANUAL
CALIBRATE SCALE
4 = Four Calibration Points, Cont.
8. With the “Known test weight” screen displayed, place the third calibrated test weight (full
test load) on the scale, enter the setting for the test weight, and then press Save.
CALIBRATE SCALE
1000.0
This is the known
test weight
that is currently
on the scale.
1 2 3
4 5 6
7 8 9
SAVE CANCEL
0 .
DEL
9. The display will change to show “Working” for a few seconds and then return to the Settings
Page 5 screen.
10. Calibration is now complete.
11. Press Exit to exit Setup and return to the Weight screen.
NOTE: A minimum of 50%
of the scale's capacity is
required for the full test
load, however 70% to 100%
is recommended.