DS5 Isolated Bipolar Constant Current Stimulator DS5 Operator's Manual 5.0
Specifications
Output: Bipolar constant current proportional to the input voltage (+ve input voltage
gives rise to an output current where RED socket becomes +ve with respect to
BLACK socket).
Output Ranges: ±10; ±25; ±50 mA for a full scale input
Input Ranges: ±1; ±2.5; ±5; ±10 V full scale
Connections: Output - 4mm shrouded, touch-proof sockets (red and black) on 3/4” centres
Monitor - Rear panel BNC sockets for “delivered current” and “applied voltage”
Input - Rear panel BNC socket
Firmware Update - USB socket on rear panel
Controls: Four front panel push buttons labelled on LCD
Output – On/Off-Reset – Momentary Action Toggle Switch
Safety Limits: 50mJ/300mJ pulse energy
50µA average “idle” current
10mA average pulse current
50mA peak current
1s/5s maximum pulse duration
Indicators: Power ON LED Green
Warning LED Amber
LCD Display for setting up and checking:
- Peak current delivered* - Input voltage range
- Output current range - Peak Voltage*
- Error warnings - Out of compliance
- Pulse energy (PE) - Pulse duration (PD)
- Average idle current (AvL) - Average pulse current (AvH)
Software: Microsoft Windows
TM
compatible DS5 User Interface (USB control)
software (Win2000 and WinXP).
Power: 100V, 120V, 200V or 240V (externally selected), 47-63Hz, 35VA
Classification: Class I with Type BF applied part
Safety: EN(IEC) 60601
Dimensions: 225mm x 100mm x 255mm (width x height x depth)
Weight: 4kg (approx.)
* These values are displayed graphically NOT numerically
* These values are displayed graphically NOT numerically
Digitimer Ltd. 17 of 52 Copyright © 2009