DS5 Isolated Bipolar Constant Current Stimulator DS5 Operator's Manual 5.0
Digitimer Limited makes no representation that the design, implementation, testing or manufacture of the
DS5 stimulator meets more than reasonable standards for use as a medical product. As stated in the
Digitimer Limited Warranty Statement, supplied with each product, “Digitimer Limited Expressly disclaims all
warranties to buyer except the limited warranty set forth therein, including without limitation to any and all
implied warranties of merchantability and fitness for a particular purpose.”
Precautions & Warnings
Operator’s Manual
Carefully study this Operator’s Manual before using the DS5 Bipolar Constant Current Stimulator.
Risks to Patient Safety
Stimulating electrodes should not be placed thoracically or trans-thoracically as this may increase
the risk of cardiac fibrillation. A warning to this effect is given on the outside of the DS5.
Care should be taken when selecting appropriate stimulating electrodes. If the current density of
the chosen electrode is greater than 2mA rms/cm
2
then the operator should pay extra attention to
stimulus parameters to ensure that the patient does not suffer burns at the stimulation site.
Simultaneous connection of a patient to the DS5 and high frequency surgical equipment may result
in burns at the site of stimulation and possible damage to the stimulator.
Operation of the DS5 in close proximity (e.g. 1m) to a shortware or microwave therapy equipment
may produce instability in the stimulator output.
The DS5 is not suitable for use in close proximity to extreme magnetic fields, such as those
generated by MRI scanners.
Patient Exclusion Criteria
Patients fitted with an implanted electronic device such as a cardiac pacemaker should NOT be subjected to
stimulation unless specialist medical opinion has been obtained.
Explosion and Fire
The DS5 must not be used in an explosive atmosphere or in the presence of flammable anaesthetic gasses.
High Voltages
Dangerous voltages are present within this unit. Do not remove covers, and refer servicing a Digitimer
representative.
Digitimer Ltd. 8 of 52 Copyright © 2009
WARNING
WARNING
For Patient and Operator Safety, this manual
should be FULLY read and understood
before use of this product