5. Measure the distance the tape advanced past the previous mark
If the tape advance distance is not within the suggested range refer to the Tape Advance Adjustment
procedure.
Micro Switch Verification:
Each CL96 analyzer uses (2) micro switches (see images below) in order to identify the position of the
gate during a tape advance. Each micro switch is held in place by (2) Philips head screws. If these screws
become loose it could affect the analyzers ability to complete a successful tape advance. Verifying
tightness of these screws can be done one time per year at the same time the gate assembly is
lubricated.