EasyManua.ls Logo

DOD CL96 - Background Verification Procedure

DOD CL96
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 2 of 44
DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300
DC-MAN-CL96Servicemanual
Innovative Life Safety Systems & Services
If there is a DAC value outside of the suggested range, verify the Chemlogic cassette is installed correctly
and press calibrate. If the DAC is still outside the suggested range refer to the CL96 Optic Cleaning
Procedure.
Background Verification:
The background value for each point can be found in the system screen by following the steps below. If
any of the background values are not between 2600 and 3300 the optics should be calibrated by
pressing the calibrate button.
1. Enter Setup menu
2. Enter Factory menu
3. Enter password 1225
4. Select the System tab
5. Press the ADC button for any active analyzer (see image below)