Operation – Part 2
20.08.2012 Page 30 of 42 Rev. 04
5.2.2.1 Calibration
By pressing the Menue key you reach the menu item “Calibration” if the
measuring method for glucose or lactate is active. In the second row of the
display, you can choose between yes” or “no” assigned to the two device
buttons. If you select “no”, the next item “Display result memory” appears; if
you press “yes”, you switch to the „Measurement Menue“ and the sample plate is
turned into the calibration position (blue). If there is a sample cup in the
calibration position, you will be asked to remove it so that the fill level can be
checked. To start calibration, insert a sample cup filled with calibration solution.
After successful calibration, the device is ready for measuring. As long as the
device is not switched back to Stand by, the device automatically checks
calibration periodically.
5.2.2.2 Display results
Here you can view the results of about the last 100 measurements. The value
measured last is displayed first; by pressing “Menue key” repeatedly the value
before that is displayed and so on. If glucose and hemoglobin
are measured
simultaneously, both results of a sample are displayed one after the other. First
the method is displayed, after a second the display changes and the result
appears in the second row. The first row displays either date and time of the
measurement or the sample identification entered. By pressing ENTER again, the
display disappears and you return to the menu item “Display result memory”.
5.2.2.3 Delete results
If the result memory is deleted, all results are lost and cannot be recovered.
Therefore, after selecting this item, you’ll be asked for confirmation “Delete
yes/no”.
5.2.2.4 Copy results
This function copies the result memory to a special data transfer card. This
allows transferring the results to a PC for further processing using a card reader.
Processing on a PC is currently under development.
5.2.2.5 Output results
Selecting this menu item and pressing ENTER outputs the measured values
stored in memory to the printer and electronic data processing system.
The results are not deleted. Results that have been output already can be read
out again by selecting “yes” after the prompt “Print all?”.
depending on model of equipment