NM2C SOFTWARE UPDATE PROCEDURE (continued)
7.6.6 Download System Software
This screen displays the list of protocols available for communication with
external monitors.
-- At the Primary Service Menu (described earlier), move the cursor to LOAD
SYSTEM SOFTWARE, and press ENTER.
MACHINE ID: NM2C
SERIAL NUMBER: 00011066
CRITICAR 1100 1.01
DATEX CARDCAP 1.00
COLIN BP-508 1.00
SIEMENS SIRECUST 1.00
CRITICAR POET IQ 1.00
* DATEX AS3 1.00
CRITIKON DINAMAP 1.00
PURITAN BENNETT 1.00
DATASCOP DIAP 1.00
DATASCOP MULTINEX 1.00
* TRAM NET 1.00
SPACELAB LOGGER 1.00
* NELLCOR N-1000 1.00
* indicates selected protocol
MAXIMUM PROTOCOL SELECTED
ESC or F10 to download code
SV50124
SOFTWARE VERSION: 1.20
SYSTEM FEATURES: 0003
Maximum protocols selected.
Figure 7-9: SELECT PROTOCOL SCREEN
-- Use the up and down arrow keys to scroll through the list; press ENTER
to select the desired protocol(s), and press ESC to initiate the downloading
process. (Ref. Paragraphs 7.4 and 7.5)
Rev. G
7-13
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS