Instructions for Use Oxylog 3000 SW 1.n 83
Configuration
1 Briefly press the corresponding button.
The associated letter on the screen changes
from "B" to "X". If the button has an LED, it will
be illuminated by the device. If there are buttons
without LED, the yellow warning LED will light
up on the device.
2 Briefly press the
O key.
The ventilator switches OFF if it is pressed for
longer than 3 seconds.
The function of the rotary knob is not included in the
test.
Test loudspeaker, buzzer, LEDs and display
To test the loudspeaker, buzzer, all LEDs and the
display:
1 Select the required test
2 Start the test. Each function is tested by the
device.
To test the screen display (Test display):
3 Turn the rotary knob; various test cards are
displayed.
The selected test remains active until the rotary
knob is pressed again.
Display accu (battery) and supply data
The parameters of the replaceable battery and the
status of the external power supply are displayed.
Display (example):
1 Advance to the second page:
2 Select the line Page, confirm and turn rotary
knob.
Display (example):
09537170
09637170f11637170

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 3000 and is the answer not in the manual?

Dräger Oxylog 3000 Specifications

General IconGeneral
Tidal Volume50 - 2000 mL
Ventilation ModesCPAP
Frequency5 - 80 bpm
Inspiratory Time0.2 - 5 s
PEEP0 - 20 cmH2O
Power Supply100 - 240 V AC, 50/60 Hz

Related product manuals