9-9
CH 9/ Downloading Events
Modem
initialization
string command
(continued)
The following is the AT initialization string for a modem using the Agere (Lucent)
chipset:
AT&F;S0=2;&K0;%E0;S30=6
Where
AT is the modem attention command.
&F instructs the modem to restore factory configuration.
S0=2 turns auto-answer on and sets the number of rings to answer to two.
&K0 turns flow control off.
%E0 disables baud rate fall back/fall forward.
S30=6 sets a 60 second inactivity modem hang up timeout.
; is a command delimiter.
A similar note regarding connect speeds applies. Try using the AT-V90= command and
modifying the contents of S registers S37 and S38
.