APPENDIX B/ Technical Specifications
Tota ls
The computations below apply to three phase wye configuration.
Totals for split phase can be determined by eliminating phase C (i.e. split phase
WATT
Tot
= WATT
A
+ WATT
B
)
Power Total
Fundamental
Power Total
Fundamental
Reactive Power
Tota l
Apparent Power,
Arithmetic Total
Apparent Power,
Vector Total
Fundamental
Apparent Power,
Arithmetic Total
Fundamental
Apparent Power,
Vector Total
Continued on next page
CBATot
WATTWATTWATTWATT ++=
fundCfundBfundAfundTot
WATTWATTWATTWATT
++=
fundCfundBfundAfundTot
VARVARVARVAR ++=
CBATotArithmetic
VAVAVAVA ++=
22
fundTotTotTotVector
VARWATTVA +=
fundCfundBfundAfundotTArithmetic
VAVAVAVA ++=
22
fundTotfundTotfundTotVector
VARWATTVA +=