4-8
Touch Screen Calibration, continued
Calibration
procedure
(continued)
Action... Result...
STEP 2: Once touch screen
calibration is done, a message will
appear to indicate Touch screen
was successfully calibrated.
Press OK to continue on with
calibration test verification.
Proceed to Step 3.
MARK167
STEP 3: Test if touch screen
recognition operates properly by
applying pressure on the LCD
screen.
Press Clear to reset LCD
screen and delete display
markings.
Press Accept to complete the
touch screen calibration
procedure. Accepting will save
and store new touch screen
calibration data in memory.
Once the new data is stored, the
old touch screen calibration
data is lost. The screen will
return to the Display Setup
Menu.
Press Cancel to discontinue the
touch screen calibration
procedure. Pressing Cancel will
retain the previous touch screen
calibration constants, no new
calibration data will be stored
in memory. The screen will
return to the Display Setup
menu.
MARK168
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dranetz PowerGuide 4400 and is the answer not in the manual?

Dranetz PowerGuide 4400 Specifications

General IconGeneral
BrandDranetz
ModelPowerGuide 4400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals