English–PACE5000/6000 Safety Instructions | 7
Figure 2: Pneumatic Connections - Typical Use Cases
2.1 Supply Equipment
Pneumatic supplies should have isolation and venting valves and, where necessary, conditioning
equipment. The positive pressure supply should be regulated to between 110% of the full-scale
pressure range and Maximum Working Pressure (MWP) stated on the control module.
To protect the control module, for ranges above 100 bar (1450 psi), from over-pressure a suitable
pr
otection device (such as a relief valve or bursting disc) must be fitted to limit the applied supply
pressure to below the Maximum Working Pressure (MWP). Refer to K0443 User Manual.
On instruments without a negative supply, the positive pressure discharges from the system to
at
mosphere through the negative supply port. Fit the diffuser to the negative port to diffuse airflow.
During system pressure vent operations, the pressure discharges from the system to atmosphere
th
rough the vent port. Fit a diffuser to the vent port to diffuse airflow.
Output Port
PACE CM
Customer
UUT
Customer Use Case 1
Customer
UUT
Customer Use Case 2
PACE CM
PACE CM
PACE CM
Output Port
Output Port
Output Port