EasyManua.ls Logo

DTS SLICE NANO - Page 33

DTS SLICE NANO
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SLICE User’s Manual support@dtsweb.com
Version 1.0 - May 2010 33 ©DTS, Inc. - All Rights Reserved
Calibration History
! This field is automatically updated whenever a new sensitivity is applied to
the sensor attributes. You cannot enter directly into this field.
Options
! Invert: When checked, the data will be inverted.
! Shunt Check and Bridge Resistance: When Shunt Check is checked, the
sensor will have the bridge resistance measured during diagnostics and
compared to the value entered in Bridge resistance.
! SW Filter: Choose the frequency of a software filter to be applied to the
data when viewing. This only affects the viewed data as all data stored will
be as collected with the hardware anti-alias filter.
! Zero Method (post download software zeroing):
o Use Pre-Cal Zero: The Zero Measured Output (ZMO) of the sensor
during calibration will be used to set the EU zero of the downloaded
data.
o Average Over Time: Used in conjunction with ZeroStart and
ZeroEnd, the average EU value during the Start and End window
will be used to Zero the collected data. The Zero Start/End window
must be set to data that will be collected. If using a negative time,
then the Acquire tab must include this window.
o None: The actual recorded input will not be adjusted or
compensated for zero level. This setting can be used to show the
actual mV offset. An example may be to record a logic level signal
and see the actual on/off state.
o Remove Offset (hardware): When checked, this will remove the
ZMO during diagnostics. This will “electrically” zero the input.
o ZeroStart/End: See Zero Method!Average Over Time.
o Offset Check: Used in conjunction with Limit Low/High during
diagnostics. When checked, the ZMO is measured and compared
the Low/High limits as a pass/fail criteria during diagnostics.
Step-by-Step Procedure to Add a New Sensor
1. On the PREPARE tab,
click the “Add” button in
the “Current Sensor”
button group

Table of Contents