11.4 SYSTEM ALERTS
Door Open - Cabinet door has been open for more than 30 seconds:
o Close door(s).
o Door switch may have become broken. Contact Support
o Door switch actuator is missing/improperly adjusted.
Dryer Failure - Dryer has been regenerated twice and has failed to reduce the humidity to within 5%
RH of the set point:
o Confirm dryer cable has not become loose or disconnected.
o Replace the dryer. (Refer to topic 10.4)
RH Probe Read Failure: The controller has failed to communicate RH probe:
o RH Probe may have become damaged or bad. (Refer to topic 10.5)
o Confirm cable has not become loose or disconnected at probe or Controller Main Board. (Refer to
topic 10.5)
SmartBAKE™ Heater Failure: The baking heater has failed to reach temperature set point within 75
minutes:
o Confirm cable has not become loose or disconnected at Power Board. (Refer to topic 10.6)