6.6.7. Pump
Displays status of pump mechanism. Units for pressure are stated in upper row.
STATUS OF PUMP MECHANISM
Item
Description
PACT
Actual pressure.
PAVG
Middle pressure during one cycle.
PULS
Pressure pulsation (bandwidth PMAX PMIN) per one cycle.
DEAD
Status of optical switch detecting position of top dead center of main (left)
piston CLOSE/OPEN.
MAGV
Motor load mean value (% of maximum allowed load).
MMAX
Motor load peak value (% of maximum allowed load).
CAMR
Camshaft speed (rotations per minute).
PPOS
Piston actual position.
6.6.8. Solvents
On main screen is displayed SOLV. In this menu item is displayed actual volume of residual
mobile phase (solvents).
6.6.9. Pulsation
On main screen is displayed PULS. In this menu item is displayed status of pump pulsation.
Units for pressure are stated in upper row.
PUMP PULSATION STATUS
Item
Description
PU01
Pulsation during last rotation.
PUAN
Average pulsation in n rotations, n is determined by firmware based on
actual flowrate.
PUBT
Best reached pulsation during learning process.
PULI
Pressure pulsation limit.
PAVG
Average pressure in one rotation.
LTIM
Learning time time lapsed from learning start.
LFLO
Learning flow flowrate during learning (constant).
PPOS
Actual piston position.
Note: After start of compensation learning (see 6.3.1) for items PUAN, PUBT displayed
---. Values of pulsation are displayed after n rotations. At this moment is learning
process started.
6.6.10. Gradient State
On main screen is displayed GRAD. In this menu item is displayed status of values for
controlling gradient valve.