Introduction HiVol 3000 User manual 1.7
4
1.2.4 Operating conditions
Ambient Temperature Range:
0-45°C
Note: Avoid corrosive, salty, or excessively
humid environments where possible.
RH 10% to 100% non condensing
1.2.5 Physical features
Case dimensions:
380x380x1200mm (WxDxH) plus inlet
Inlet materials:
Anodised aluminium and stainless steel
Interchangeable Inlets:
TSP, PM
10
and PM
2.5
(options)
Cut-point:
PM
10
and PM
2.5
50% cut-point at 67.8m
3
/h volumetric flow-rate
Filter holder material:
Anodised aluminium and stainless steel
Filter element:
250x200mm rectangular element (various materials and types are available)
Weight:
45 kg plus inlet
1.3 Nomenclature
SSI: Size Selective Inlet, used to sample particulates of a given Equivalent
Aerodynamic Diameter (PM
10
, PM
2.5
).
TSP: The TSP (Total Suspended Particulate) includes all suspended
particulate matter.
PM
10
: The PM
10
includes all particulates of an Equivalent Aerodynamic
Diameter (EAD) of 10 microns or less.
PM
2.5
: The PM
2.5
includes all particulates of an Equivalent Aerodynamic
Diameter (EAD) of 2.5 microns or less.
PPM: Parts Per Million