HiVol 3000 User manual 1.7 Introduction
9
1.4.4.2 Australian Ambient Particulate Standards
The Australian National Environmental Protection Council has established
the following objectives for Particles as PM
10
:
24 hour average
Maximum concentration: 50µg/std m
3
Maximum allowable excedances: 5 days per year
Australian reference conditions:
Standard temperature defined as 273K (0°C)
Standard pressure defined as 101.3 kPa (760mm Hg)
1.4.4.3 U.S. EPA reference method designation
For the HiVol to be used as a U.S. EPA designated reference method the
sampler must be configured in the following arrangement:
• PM
10
size selective inlet installed
• Flow rate set to 67.8 m
3
/hour (1.13m
3
/min)
Any configuration different to this is not covered under the designation.