Front Matter HiVol 3000 User Manual 1.7
vi
Applicable to optional CE Version HiVol 3000 only
Electrical and Safety Conformity.
Ecotech P/L certifies that this product operates in compliance with the EC Directive
89/336/EEC in reference to electrical emissions and immunity. Specifically the
equipment meets the requirements of EN61000-6-3 for RF Emissions and EN 61000-
6-1 for RF immunity.
In addition, this product has been tested in compliance with the EC Directive
73/23/EEC in reference to electrical safety and meets the requirements of EN 61010-
1.
The equipment must be operated as per Ecotech directions as specified in this manual.
Safety Requirements
To reduce risk of personal injury caused by electrical shock, follow all safety
notices and warnings in this documentation.
This equipment should always be used with a protective earth installed.
The HiVol 3000 is compliant with the requirements of EN 61010-1, Safety
Requirements for Equipment for Measurement, Control, and Laboratory Use.
If the equipment is used for purposes not specified by the manufacturer, the
protection provided by this equipment may be impaired.
Replacement of any part should only be carried out by qualified personnel,
only using parts specified by the manufacturer. Always disconnect power
source before removing or replacing any components.
Equipment Rating
240V~ ±10% (not applicable to 110V version)
50/60 Hz
1500 VA max
All wiring must be in accordance with local norms and be carried out by
experienced personnel.
IP56 rated
Fuse: 8 Amp T250
Environmental Conditions
Relative humidity 10% to 100% non
condensing
Temperature
0 to 45 °C
Pollution degree 2
Installation category II
Maximum altitude 2000m.
Never operate this equipment in the presence of flammable liquids or
vapours, as this could cause a safety hazard.