Site meets the relevant siting criteria for jurisdiction in which monitoring
are to be performed.
Electrical connections and earth leakage circuit breaker have been
installed and checked.
Sampler has been securely fastened to supporting structure.
When using the optional SSI, the impaction shim has been greased and
installed in the size selective inlet.
Field calibration has been performed and flow rate meets the inlet design
flow rate requirement.
A pre-weighed filter has been installed and details including the initial
flow rate have been recorded in an appropriate log sheet.
STP reference conditions have been entered in the Set-up Menu.
The sampler timer has been programmed for the required run-time and
sampling frequency.