• Site meets the relevant siting criteria for jurisdiction in which monitoring
are to be performed.
• Electrical connections and earth leakage circuit breaker have been
installed and checked.
• Sampler has been securely fastened to supporting structure.
• When using the optional SSI, the impaction shim has been greased and
installed in the size selective inlet.
• Field calibration has been performed and flow rate meets the inlet design
flow rate requirement.
• A pre-weighed filter has been installed and details including the initial
flow rate have been recorded in an appropriate log sheet.
• STP reference conditions have been entered in the Set-up Menu.
• The sampler timer has been programmed for the required run-time and
sampling frequency.