Cadence II Fetal Monitor Service Manual
56
Although transducers are designed for durability, they should be handled with care.
Rough handling could damage the cover, piezoelectric crystals and mechanical
movement. The cover is made of a soft plastic, and contact with hard or sharp objects
should be avoided. Do not excessively flex the cables.
WARNING : Under no circumstance are transducers to be used to monitor patients
under water.
CAUTION :
Be sure that the cleaning solutions and transducers do not exceed a temperature of
45.
Do not autoclave the transducers and cables or heat them above 70.
6.3.1 Cleaning of TOCO Transducer
To clean a TOCO transducer:
Wipe the transducer with a cloth.
Clean the transducer with a cloth soaked in a solution of soap and water, or a
cleaning solution. Do not immerse the transducer in the solution. Use only the
following cleaning solutions:
Z BURATON LIQUID
Z MIKROZID
Z ETHANOL 70%
Z SPORACIDIN
Z CIDEX
When using a cleaning solution, follow the manufacturers directions carefully to
avoid damaging the transducer.
Wipe the transducer with a cloth damped with water.
Wipe the transducer with a clean, dry cloth to remove any remaining moisture.
6.3.2 Cleaning of Ultrasound Transducer
To clean an ultrasound transducer: