Cadence II Fetal Monitor Service Manual
57
Wipe the transducer with a cloth.
Immerse the transducer in a solution of soap and water, or a cleaning solution. Use
only the following cleaning solutions:
Z BURATON LIQUID
Z MIKROZID
Z ETHANOL 70%
Z SPORACIDIN
Z CIDEX
When using a cleaning solution, follow the manufacturers directions carefully to
avoid damaging the transducer.
Rinse the transducer with water.
Wipe the transducer with a clean, dry cloth to remove any remaining moisture.
6.3.3 Cleaning of Fetal ECG Cable
The fetal ECG cables must never be immersed, soaked, or cleaned with harsh
chemicals. The recommended cleaning method of the cables is a cloth wipe using
ordinary alcohol-free hand soap or USP green soap tincture. After cleaning the cables,
they should be wiped with water using a clean damp cloth and then with a clean dry
cloth.
Do not autoclave, Ethylene Oxide (EtO), or stem sterilize the fetal ECG cables.
6.4 Cleaning of Recorder
The recorder platen, thermal print head and paper sensing mechanism must be cleaned
at least once a year or when needed (when traces become faint).
To do this:
Clean the recorder platen with a lint-free cloth and soap/ water solution.
With the thermal array using a cotton swab moistened with 70% Isopropyl
alcohol-based solution.
Check the paper sensing mechanism is free of dust.
6.5 Cleaning of Belt
Wash soiled belts with soap and water. The water temperature must not exceed 60.