EasyManua.ls Logo

EDAN DUS 60 - Page 14

EDAN DUS 60
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 9 -
Acoustic Output Reporting Table
Transducer Mode C321UA
Operation Mode B+M Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.8458 0.1768 0.0178 0.0639 0.5391
P
r.3
MPa 1.506
W
0
mW 31.4 2.317 33.717
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
1.23
Z
1
(cm) 3.0
Z
bp
(cm) 2.3417
Z
sp
(cm) 4.45
Z@PII
.3max
(cm) 4.45
deq(Z
sp
) (cm) 0.3579
f
c
(MHz) 3.17 3.17 3.17 3.17 3.17
X(cm) 1.28 1.28 1.28 1.28
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
1.5 1.5 1.5 1.5
PD (usec) 0.4958
PRF (Hz) 1518
P
r
@PII
max
(MPa) 2.4931
d
eq
@PII
max
(cm) 0.3579
FL
x
(cm) 2.186 2.186 2.186 Focal
Length
Fl
y
(cm) 42.2692 42.2692 42.2692
Other
Information
I
pa.3
@MI
max
(W/cm
2
)
Control
Conditions
Control1
AP=15; Frequency=3.0MHz;
Depth=98mm; Focus=60mm;

Table of Contents

Other manuals for EDAN DUS 60

Related product manuals