DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 10 -
Acoustic Output Reporting Table
Transducer Mode C321UA
Operation Mode PW Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.4616 0.6412 3.0623 1.3986
P
r.3
MPa 0.8082
W
0
mW 87.47 87.47
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
50.0
Z
1
(cm) 3.0
Z
bp
(cm) 2.3417
Z
sp
(cm) 4.95
Z@PII
.3max
(cm) 4.95
deq(Z
sp
) (cm) 0.0075
f
c
(MHz) 3.065 3.065 3.065 3.065
X(cm) 1.28 1.28 1.28
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
1.5 1.5 1.5
PD (usec) 1.9784
PRF (Hz) 6361
P
r
@PII
max
(MPa) 1.2714
d
eq
@PII
max
(cm) 0.0075
FL
x
(cm) 2.6236 2.6236Focal
Length
Fl
y
(cm) 54.189 54.189
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.036
Control
Conditions
Control1
AP=15; Frequency=2.0MHz;
Depth=98mm; Focus=60mm;