DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 21 -
Acoustic Output Reporting Table
Transducer Mode C362UA
Operation Mode B+M Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
≤1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.8549 0.0922 0.0146 0.0582 0.4652
P
r.3
MPa 1.528
W
0
mW 30.57 1.898 32.468
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
1.01
Z
1
(cm) 30.0
Z
bp
(cm) 2.6132
Z
sp
(cm) 4.45
Z@PII
.3max
(cm) 4.45
deq(Z
sp
) (cm) 0.3908
f
c
(MHz) 3.2169 3.2169 3.2169 3.2169 3.2169
X(cm) 1.5936 1.5936 1.5936 1.5936
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
1.5 1.5 1.5 1.5
PD (usec) 0.4834
PRF (Hz) 1836
P
r
@PII
max
(MPa) 2.3303
d
eq
@PII
max
(cm) 0.0243
FL
x
(cm) 1.7369 1.7369 1.7369Focal
Length
Fl
y
(cm) 26.6692 26.6692 26.6692
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.1409
Control
Conditions
Control1
AP=15; Frequency=2.0MHz;
Depth=98mm; Focus=60mm;