DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 22 -
Acoustic Output Reporting Table
Transducer Mode C362UA
Operation Mode PW Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.4772 0.5209 2.1213 0.9647
P
r.3
MPa 0.7812
W
0
mW 67.33 67.33
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
35.5
Z
1
(cm) 30.0
Z
bp
(cm) 2.6132
Z
sp
(cm) 4.45
Z@PII
.3max
(cm) 4.45
deq(Z
sp
) (cm) 0.0108
f
c
(MHz) 2.6875 2.6875 2.6875 2.6875
X(cm) 1.5936 1.5936 1.5936
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
1.5 1.5 1.5
PD (usec) 1.4124
PRF (Hz) 8186
P
r
@PII
max
(MPa) 1.3549
d
eq
@PII
max
(cm) 0.0108
FL
x
(cm) 1.9452 1.9452Focal
Length
Fl
y
(cm) 33.9865 33.9865
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.0338
Control
Conditions
Control1
AP=15; Frequency=3.0MHz;
Depth=98mm; Focus=60mm;