DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 30 -
Acoustic Output Reporting Table
Transducer Mode C363UA
Operation Mode B+M Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.8507 0.1166 0.0182 0.0598 0.4682
P
r.3
MPa 1.493
W
0
mW 35.2 2.413 37.613
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
1.3
Z
1
(cm) 3.008
Z
bp
(cm) 3.008
Z
sp
(cm) 5.45
Z@PII
.3max
(cm) 4.5
deq(Z
sp
) (cm) 0.3944
f
c
(MHz) 3.083 3.083 3.083 3.083 3.083
X(cm) 2.112 2.112 2.112 2.112
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
1.5 1.5 1.5 1.5
PD (usec) 0.504
PRF (Hz) 1634
P
r
@PII
max
(MPa) 2.3798
d
eq
@PII
max
(cm) 0.3944
FL
x
(cm) 1.5429 1.5429 1.5429Focal
Length
Fl
y
(cm) 20.3404 20.3404 20.3404
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.1041
Control
Conditions
Control1
AP=15; Frequency=2.0MHz;
Depth=98mm; Focus=70mm;