DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 31 -
Acoustic Output Reporting Table
Transducer Mode C363UA
Operation Mode PW Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.5314 0.5852 2.75 1.1357
P
r.3
MPa 0.8781
W
0
mW 79.26 79.26
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
45.01
Z
1
(cm) 3.0
Z
bp
(cm) 2.6132
Z
sp
(cm) 5.0
Z@PII
.3max
(cm) 5.0
deq(Z
sp
) (cm) 0.0083
f
c
(MHz) 2.7296 2.7296 2.7296 2.7296
X(cm) 2.112 2.112 2.112
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
1.5 1.5 1.5
PD (usec) 1.4049
PRF (Hz) 6361
P
r
@PII
max
(MPa) 1.4072
d
eq
@PII
max
(cm) 0.0083
FL
x
(cm) 1.6648 1.6648Focal
Length
Fl
y
(cm) 22.1698 22.1698
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.0441
Control
Conditions
Control1
AP=15; Frequency=2.0MHz;
Depth=98mm; Focus=70mm;

Table of Contents