SE-12 Series Electrocardiograph User Manual EMC Information
- 104 -
Recommended separation distances between portable and mobile RF communications
equipment and the EQUIPMENT or SYSTEM
Recommended separation distances between
portable and mobile RF communications equipment and the 12-channel Electrocardiograph
The 12-channel electrocardiograph is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the 12-channel
electrocardiograph can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
12-channel electrocardiograph as recommended below, according to the maximum output power
of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter(m)
150 kHz to 80 MHz
Pd 2.1
80 MHz to 800 MHz
Pd 2.1
800 MHz to 2.7 GHz
Pd 3.2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.