SE-12 Series Electrocardiograph User Manual Transmitting ECG Data
- 44 -
4) Set the FTP User Name, FTP Password and FTP Path items.
a) The user name and the password you input in the FTP User Name and FTP
Password items must be available for FTP server.
b) The path you input in the FTP Path item must be the subdirectory of the path you
input in the FTP receiving software.
NOTE: For more information about FTP server, consult your Network Administrator.
3. Set File Format in the File Setup window.
4. In the auto or rhythm mode, ECG data will be transmitted through the network automatically
after an ECG report is printed out.
NOTE: If the message Transmitting Fails appears on the system screen, then there has
been an error in network communication of the data. Please check the network
connection, move the device where the wireless signal is stronger or directly use
a U disk to transfer data.
7.2 Real-time Transmission to the DMS
In real-time transmission, the electrocardiograph functions as an ECG sampling box to the DMS
software.
NOTE: To use the real-time transmission function, the ECG data management software
(DMS) of EDAN must be installed in the PC.
1. Start the DMS.
2. In the Sampling Setting window, set Device Model to SE12.
3. Connect USB socket 1 of the electrocardiograph to the USB socket of the PC by using the
high-speed USB cable.
For details, please contact the manufacturer or the local distributor.
4. Start real-time transmission
The sampling box transmits to the PC the ECG signals acquired from the patient. Acquisition
and transmission are simultaneous. The ECG signals are displayed on the PC monitor and
eventually analyzed. For more details, refer to the user manual of the DMS software.